EcoSEC Elite - Software Specifications

Data Acquisition

Data AcquisitionSpecifications
Input2-channel (RI, UV or external detector) / I/O connection necessary
Acquisition time0.0 to 300.0 minutes
Acquisition interval10 ms or more (10 ms steps). Upper limit: 1,000 ms

 

Calibration

CalibrationSpecifications
Calibration curve approximationFirst-degree expression
3rd-degree expression
3rd-degree expression + hyperbola
5th-degree expression
7th-degree expression
7th-degree expression (odd power)
7th-degree expression (odd power) + hyperbola
Calibration curve correction Mark-Houwink
Q factor
Polymerization degree
USP

 

Calculations

CalculationsSpecifications
Quantitative calculation specific to GPCMolar mass averages (Mn, Mw, and Mz )
Polydispersity Index (PDI)
Cumulative/differential molar mass distributions
Concentration ratio
Special calculation functionInternal standard correction function
Copolymer analysis
Molar mass fraction specific calculation
Calculation range specification
Lag time correction
Calculation standardASTM®
DIN®
USP
JIS
JP
ISO 16014
Tosoh Standard

 

Control

ControlSpecifications
Warm up and shut down timersDaily
Weekly
RI and UV detector auto balanceOptional prior to injection

 

Column Test

Column TestSpecifications
ParameterTheoretical plate number
Resolution
Symmetry factor
Half bandwidth

 

FDA CFR 21 Part 11

FDA 21 CFR Part 11Specifications
FunctionalitySoftware validation, authentication by user ID and password, log out, and audit trail