EcoSEC™ High Temperature System
Demanding high temperature polymer analyses, such as for polyolefins, require a system that delivers results reliably, reproducibly, and produces these results in an easy to use, safe instrument. The EcoSEC High Temperature GPC System incorporates the proven design and technology used in our ambient EcoSEC Elite™ GPC System.
The dual pump system, dual flow RI detector, spacious column oven, total system temperature control, autosampler, sample prep station, safety features, and all-in-one design controlled by intuitive software combine to deliver exceptional performance.
The EcoSEC High Temperature GPC System can be upgraded with a High Temperature Viscometer Detector.
| The EcoSEC High Temperature GPC System features: |
|---|
| Reliability |
| Reproducibility |
| Stability |
| Safety |
| All-in-one design |
| Ease of use |
Features & Benefits
| What is unique about the EcoSEC High Temperature GPC System? | |
|---|---|
| Dedicated system that is easy to use and is ready when you are » | Reliability
Baseline Stability
Reproducibility
Easy Operation
Global Standards
|
| Dual flow RI detector, delivering unparalleled RI baseline stability » | Learn more |
| Total system temperature control, resulting in accurate flow rates and reproducible molar mass determinations with consistent results from run-to-run, day-to-day, and system-to-system » | The EcoSEC High Temperature GPC System has four independent heating elements that maintain a constant temperature within the range of 40 ºC to 220 ºC: pump oven, column oven, autosampler, and RI detector. Maintenance of a constant temperature of the pump heads and solvent lines of the EcoSEC High Temperature GPC System allow the pumps to deliver exactly the same flow rate under any mobile phase condition, thus improving baseline stability. Most HPLC pumps do not compensate for volume changes as a function of temperature. The autosampler in the EcoSEC High Temperature GPC System is temperature controlled by aluminum block and is independent of the pump and column oven. A stable baseline results from the temperature controlled RI detector. |
Components

| Click on the individual component links below for more detailed information | |
|---|---|
| Solvent Reservoir » | The solvent reservoir on the EcoSEC High Temperature GPC System maintains solvents at a constant temperature of 40 °C to prevent possible solvent freezing. The capacity of the solvent holder is 1.4 L bottle. |
| Control Panel » |
|
| Temperature Controlled Pumps » | The EcoSEC High Temperature GPC System has four independent heating elements that maintain a constant temperature within the range of 40 ºC to 220 ºC: pump oven, column oven, autosampler, and RI detector. Maintenance of a constant temperature of the pump heads and solvent lines of the EcoSEC High Temperature GPC System allow the pumps to deliver exactly the same flow rate under any mobile phase condition, thus improving baseline stability. Most HPLC pumps do not compensate for volume changes as a function of temperature. The autosampler in the EcoSEC High Temperature GPC System is temperature controlled by aluminum block and is independent of the pump and column oven. A stable baseline results from the temperature controlled RI detector. |
| Column Oven » |
|
| Autosampler » |
|
| RI Detector » | Learn more |
| Purge Unit and Degasser » | The purge unit and degasser in the EcoSEC High Temperature GPC System allows rapid solvent changes via a purge valve, thus saving time by eliminating solvent replacement and other manual operations. |
| EcoSEC HT Workstation Software » | The EcoSEC High Temperature GPC Workstation includes the software, which offers a fully functional interface for instrument control, data collection, and report generation of your data. |
| Column Switching Valve Option » | Learn more |
| Sample Prep System Option » | The Sample Prep System is an optional accessory to the EcoSEC High Temperature GPC system. It dissolves and filtrates up to 24 samples, improving work flow efficiency. The Sample Prep System has a temperature range of 40 °C to 220 °C. |
Tosoh Bioscience can tailor a system to meet your application needs.
Does your analysis require additional detectors beyond RI?
The EcoSEC High Temperature GPC System provides easy, effortless, and temperature-controlled connectivity for multi-detector configurations. We offer external light scattering and viscometry detectors, as well as coupling options for FT-IR analysis.
Contact us for a quote!
Specifications
| Instrument | |
|---|---|
| Dimension | 1,000 (W) × 500 (D) × 650 (H) mm |
| Auto-injector | |
|---|---|
| Injection volume | 10 to 500 µL (dependent on sample loop size) |
| Number of samples | 24 |
| Column oven | |
|---|---|
| Temperature range | 40 °C to 220 °C |
| Capacity | 7.8 mm ID × 30 cm × 8 columns |
| Pump | |
|---|---|
| Flow rate | 0.1 to 2.0 mL/min in 0.1 mL/min steps |
| Accuracy | +/- 2% |
| Precision | +/- 0.2% |
| Max. Pressure | 15 MPA or 2,175 psi |
| RI Detector | |
|---|---|
| Type | Bryce or dual flow type. Tungsten light source |
| Cell volume | 10 µL |
| Noise | 1.5 × 10-8 RI units (RIU) (ODCB: 1.0 mL/min, 145 ºC, response: 3 s) |
| Drift | 3 × 10-7 RIU/h (ODCB: 1.0 mL/min, 145 ºC) |
Software
| The EcoSEC High Temperature GPC System Workstation Software offers unparalleled versatility and includes these features: |
|---|
| GPC-specific EcoSEC High Temperature GPC System software to simplify system control and data handling |
| Controls up to 2 EcoSEC High Temperature GPC Systems |
| Excellent data handling and report generation |
| Fully featured data handling system; analyze data from two detectors |
| Start and stop system automatically |
| One license for multiple locations |
| Software features include: | |
|---|---|
| Flow Diagram » | The flow diagram is a useful tool to modify running conditions of an individual component. |
| Method Templates » |
|
| Peak Editing » |
|
| Data Management » |
|
| Multiprocessing Function » |
|
| GPC Specific Quantitative Calculations » |
|
| Report Generation » |
|
Calibration Standards
TSKgel® polystyrene bulk calibration standards are used to calibrate GPC columns for subsequent analysis of unknown samples. The standards range from 400 to 2.1 × 107 Da.

TSKgel GPC Columns
TSKgel H series GPC US | EU columns are recommended for the analysis of organic-soluble polymers and are packed with spherical particles composed of polystyrene cross-linked with divinylbenzene (PS-DVB). The TSKgel HT columns are for high temperature applications (≤ 140 °C) while the TSKgel HT2 columns are used in ultra-high temperature (up to 220 °C) applications.
The figure below demonstrates the performance stability of the TSKgel GMHHR-H HT columns compared to other commercially available high temperature GPC columns during repetitive temperature changes. The TSKgel HHR HT columns and two commercially available high temperature GPC columns were subjected to drastic changes in temperature by raising the temperature for 2 hours followed by lowering the temperature for two hours for a total of 60 cycles. The number of theoretical plates was shown to remain constant for the TSKgel HHR HT columns and to decrease by 15% for the two commercially available high temperature GPC columns; thus revealing the superior performance stability of the TSKgel HHR HT columns.
Column Durability

| Column: | TSKgel GMHHR-H HT2, 5 µm, 7.8 mm ID × 30 cm × 2 |
| Mobile Phase: | ODCB with 0.05% BHT |
| Flow rate: | 1.0 mL/min |
| Detection: | RI (EcoSEC High Temperature GPC System) |
| Temperature: | 40 to 145°C |