Skip to main content
Selected AIA 2000 Analyser
AIA 2000 Analyser

Description

High powered flexible immunoassay analyzer with a throughput of 200 tests/hour. The AIA-2000 is a random access analyzer and uses the AIA-PACK series of assays. The LA model is compatible with most total laboratory automation tracklines.

Specifications

EU Certification
No
Sample containers
Test tubes :13x75, 13x100, 16x75, 16x100 mm (Diameter x Length). Can be used in combination with dedicated sample cups
CISPR class and group categories
Class A, Group 1
Specimen barcodes
CODE39, CODE128, ITF, NW-7, JAN compatible (compliant with ASTM14466-92 and CLA guideline compliant barcodes) Min. element width: 0.191mm (0.254mm or over recommended)
Measuring conditions
Reaction temperature: 37 degrees Celsius. Antigen-antibody reaction time: 10 min., 20 min, or 40 min. (depending on assay specification)
Barcode Reader
Class 1: IEC60825-1: 2001
Dust
Average office level
FCC
Part 15, Sub Part B, Class A
Power Supply/Power Consumption
AC 100 - 240 V, 50/60 Hz. 700 VA (excluding controller PC and printer (option)) Over voltage category II
Specimen pretreatment.
Automatic pretreatment (10 min. at 37 degrees Celsius)
Specimen dilution factors
2 per assay specification selectable in the 2 to 1000 range.
Sample loader
200 specimens, (add/change as required) STAT: Using the Priority rack
Detection method.
Fluorescent measurement (top-to-top system)
In vitro diagnostic reagent
AIA-PACK Series
Sample volumes (uL)
10 to 125 μL
External Input/Output
RS-232C (ASTM) or Ethernet (HL7 V2.4)
Sample clot detection
Pressure detection
Detection Technology
Fluorometric enzyme immunoassay (FEIA)
Processing method
Automated continuous random access
Tip Sorter
Dedicated tips. 576 (96/rack, in 6 racks). Can be added or changed at any time by key request
Pollution degree
2
Precision control
Multirule (Levey-Jenning’s charts)
EMC standard
IEC61326-1: 2005 / EN61326-1: 2006, IEC61326-2-6: 2005 / EN61326-2-6: 2007
External Dimensions/Weight (excluding controller PC and monitor)
1500(W) x 1197(D) x 1260(H) mm / 400 kg 59.1(W) x 47.1(D) x 49.6(H) inches / 881.8 lb
Temperature
15 to 30 degrees Celsius
Humidity
40% to 80% R.H. (with no condensation)
Throughput (test/h)
Max. 200 tests/h. (ST reagents)
Reagent Sorter
Max. 48 trays (equivalent to 48 analytes or 960 tests) Can be added or changed at any time by key request

Recently Viewed Items

Loading recently viewed products...
Loading product details...