The AIA-CL1200 Automated Immuno Analyzer is a high-performance| fully automated immunoassay analyzer designed for clinical laboratories requiring medium to high throughput testing.
Part #: 0024130
Brand: Tosoh
Manufacturers:
| EU Certification | No |
|---|---|
| Sample containers | Test tubes :13x75, 13x100, 16x75, 16x100 mm (Diameter x Length). Can be used in combination with dedicated sample cups |
| EMC standard | IEC61326-1: 2012, IEC61326-2-6: 2012 |
| CISPR class and group categories | CIPR 11 Class A, Group 1 |
| Specimen barcodes | CODE39, CODE128, ITF, NW-7, JAN compatible (compliant with ASTM14466-92 and CLA guideline compliant barcodes) Min. element width: 0.191mm (0.254mm or over recommended) |
| Measuring conditions | Reaction temperature: 37 degrees Celsius. Antigen-antibody reaction time: 10 min., 20 min, or 40 min. (depending on assay specification) |
| Dust | Average office level |
| FCC | Part 15, Sub Part B, Class A |
| Tip Sorter | Dedicated tips. 480 (96/rack, in 5 racks). Can be added or changed at any time by key request |
| Power Supply/Power Consumption | AC 100 - 240 V, 50/60 Hz. 1500 VA (excluding controller PC and printer (option)) Over voltage category II |
| Throughput (test/h) | Max. 120 test/h. |
| Specimen pretreatment. | Automatic pretreatment (10 min. at 37 degrees Celsius) |
| External Dimensions/Weight (excluding controller PC and monitor) | 1310(W) x 840(D) x 1360(H)mm / 477 kg |
| Detection method. | Chemiluminescent Envzyme Immunoassay (CLEIA) |
| Altitude | Up to 2000 m |
| Barcode Reader | Class 2 Laser Product: IEC60950-1:2005+A1:2009 |
| External Input/Output | RS-232C (ASTM) or Ethernet (HL7 V2.4) |
| Sample clot detection | Pressure detection |
| Detection Technology | Immuno Chemilumenisence assay (CLA) |
| In vitro diagnostic reagent | AIA-CL PACK Series |
| Sample loader | 100 specimens, (5 specimens/rack x 20 racks) (add/change as required) STAT: Using the Priority rack |
| Processing method | Automated continuous random access |
| Pollution degree | 2 |
| Specimen dilution factors | 2 per assay specification selectable in the 4 to 625 range. |
| Precision control | Multirule (Levey-Jenning’s charts) |
| Temperature | 15 to 30 degrees Celsius |
| Humidity | 40% to 80% R.H. (with no condensation) |
| Sample volumes (uL) | 5 to 50 µl |
| Reagent Sorter | Max. 640 cups (32 cups/tray x 20 trays) Can de added or changed at any time by key request |