Skip to main content
Selected Diagnostics-IND- Product- AIA-Troponin I
Diagnostics-IND- Product- AIA-Troponin I

Description

Troponin I (TnI) has three isoforms: cardiac, fast skeletal and slow skeletal and out of them cardiac troponin I (cTnI) is highly specific to the myocardium. When the myocardium is damaged, cTnI is released into the blood. The level of cTnI becomes abnormally high 3-4 hours after the onset of acute myocardial infarction and reaches the peak 10-16 hours after the onset. The level of cTnI remains abnormally high for 5-8 days after the onset. When compared to Myoglobin or CK-MB, the degree of its change in relation to the cut-off value is much greater for cTnI (undetectable in healthy individuals) and as a result, it is possible to detect minute myocardial lesions. In addition, increased cTnI levels in heart failure patients indicates latent cardiomyopathy.

Technical specifications

Kit nameST AIA-PACK cTnI 3rd-Gen 
MethodEnzyme Immunoassay (EIA)
Assay principleTwo-site immunoenzymometric assay
Assay sampleSerum, heparinized plasma or EDTA plasma
Calibrators0, 0.2, 1.0, 5.0, 20, 60 ng/mL
Standardization Internal
Calibration stability90 days
Assay range0.02 - 50 ng/mL 
Reaction time10 minutes
Hook effect>20,000 ng/mL
Storage2 - 8oC

 

 

 

 

 

 

 

 

 

 

 

 

Refer to the product Instruction-For-Use (IFU) for complete details.